Product Specifications
MINIMUM PC SYSTEM REQUIREMENTS:
MEDICAL SYSTEM (ECG, HOLTER, NIBP) FEATURES:
The cuff is long enough to completely encircle the largest circumference limb (44cm), and maintains its full width throughout this length.
Pressure Capacity: The cuff completely retains the bladder when the bladder is inflated to a minimum pressure of 330mmHg.
The tubes and connectors remain intact at 330mmHg.
Cuff Closures/construction for Non-disposable Cuffs: The cuff integrity is maintained after 1000 open-close cycles, and after 10,000 cycles to 300mmHg.
The cuff bladder length encircles much of the circumference of the limb throughout its length and maintains it’s much of its width throughout its length.
The bladder length is at least 0.80 times the circumference of the limb at the midpoint of the cuff application. The cuff bladder width is at least 0.37 (preferably 0.40) times the circumference of the limb at the midpoint of the cuff application.
System leakage (cuff bladder, tubing, and connectors, etc) is less than 1 mmHg per second.
MEDICAL ELECTRICAL PERFORMANCE AND SAFETY:
GENERAL SYSTEM:
Note: the instrument's specifications may degrade if operated outside of these environmental specifications
CASSETTE PLAYBACK SPECIFICATIONS:
SPO2 SPECIFICATIONS:
TEMPERATURE SPECIFICATIONS:
WEIGHT SPECIFICATIONS:
HOLTER FLASH RECORDER SPECIFICATIONS:
ELECTRICAL:
- IBM compatible Pentium or higher
- 512MB ram for the program for Windows 98, 2000, XP, Vista 32 and 64 bit, Windows 7, Windows 8 and Mac’s with Boot Camp, Dual Boot, VMware Fusion or Parallels. CD ROM
- USB Ports
- 5 MB hard drive space for programs and additional for patient data. For standard 12- lead tests only 60kb are needed for Holter,
- 33MB are needed per 24-hour patient test
- Windows 98, 2000, XP, Vista 32 and 64 bit, Windows 7, Windows 8 and Mac’s with Boot Camp, Dual Boot, VMware Fusion or Parallels
- Display: Any Windows compatible display
- Printers: Any Windows compatible printer
- Mouse: Any Windows compatible mouse
- Keyboard: Any Windows compatible keyboard
MEDICAL SYSTEM (ECG, HOLTER, NIBP) FEATURES:
- Leads: 12- lead systems: 12 (+ XYZ optional) on PCC16. Standard 12 with BOX12
- Holter Systems: 3 to 12- leads
- Display: LCD, color, monochrome
- ECG storage to any device that appears to the pc as a drive. e.g. hard drives, floppies, optical, etc
- Data Stored: ECG waveforms, time and date, and patient information (i.e. demographics)
- ECG transmission by direct connect, modem, fax and internet
- Measurement's Tolerances: time: +/- 4msec
amplitude: +/- 1.22uV at 10mm/mV gain - Holter Analysis: Standard arrhythmia classification of all beats detected
- NIBP's Tolerances: time: +/- 4msec
pressures: +/- 2mmHg
pulse rate: +/- 1bpm - NIBP Limits: 45 - 300 mmHg (system cut-off at 300mmHg)
- NIBP Cuff, Tubes, and Connectors: Cuff Length: For Bandage Cuffs: The full cuff length extends beyond the end of the inflatable bladder by at least the equivalent of the length of the bladder.
The cuff is long enough to completely encircle the largest circumference limb (44cm), and maintains its full width throughout this length.
Pressure Capacity: The cuff completely retains the bladder when the bladder is inflated to a minimum pressure of 330mmHg.
The tubes and connectors remain intact at 330mmHg.
Cuff Closures/construction for Non-disposable Cuffs: The cuff integrity is maintained after 1000 open-close cycles, and after 10,000 cycles to 300mmHg.
The cuff bladder length encircles much of the circumference of the limb throughout its length and maintains it’s much of its width throughout its length.
The bladder length is at least 0.80 times the circumference of the limb at the midpoint of the cuff application. The cuff bladder width is at least 0.37 (preferably 0.40) times the circumference of the limb at the midpoint of the cuff application.
System leakage (cuff bladder, tubing, and connectors, etc) is less than 1 mmHg per second.
- Cables: All electrical cables have individually insulated wires and a jacket (usually PVC). All cables can withstand a stretch force of up to 5 lbs
- Connectors: All electrical connectors have at least 0,025" spacing between conductors, have a maximum conductor contact resistance of 10ohms, and maintain this maximum resistance after 100 connect-disconnect cycles
- NIBP Features: Report: Systolic pressure, Diastolic pressure, heart rate, and in full disclosure pulse and pressure waveforms
- Indications: Ages Infant - Adult Arm Circum 8-44cm
- Contraindications: Open wounds to limb distal to cuff application
MEDICAL ELECTRICAL PERFORMANCE AND SAFETY:
- Frequency Response: .05 - 120 Hz. upper 3db at 120 Hz (AAMI & AHA Standards)
- Lead Leakage: < 10 uA (AAMI Standards)
- Chassis Leakage: < 100 uA (AAMI Standards)
- Input Impedance: > 100 M (AAMI Standards)
- Gain Sensitivity: 5, 10, 10/5 and 20 mm/mV (AAMI)
- CMRR: 120 db (AAMI)
- 12- lead ECG Sampling Rate: 250 - 1000 Hz
- A/D Resolution: 13 bits
- A/D Rate: 10 kHz
- Time Base: 25 & 50 mm/sec (AAMI)
GENERAL SYSTEM:
- Power Consumption: < 300 mA from the pc bus
- CC-Holter Recorder: 1 AA battery for 24/48 hours of continuous use
- Dimensions:
PCC Rev B: 5.5" x 4.5" x 0.6"
PCC Rev L: 6.5" x 4.5" x 0.6"
Box12 Rev M: 4.5" x 2.5" x 0.8"
CC-Holter Rev A: 6.25" x 2.75" x 0.75"
Flash Holter Recorder: 4.4" x 2.6" x .95"
ISOT Rev B: 6" x 5" x 4.25"
CCNIBP-I: 5.75" x 3.6" x 1.75" - Weights:
PCC Rev B: 118g (4.2oz) PCC Rev L: 118g (4.2oz)
Box12 Rev M: 268g (9.5oz)
ISOT Rev B: 1806g (4lbs)
CC-Holter Rev A: 268g (9.5oz) + 1 AA battery
Flash Holter Recorder: 5oz with battery
CCNIBP-I: 350g (12.5oz) - Environmental Max Temperature Range: 10°C (50°F) to 40°C (104°F)
- Recommended Range: 15°C (59°F) to 35°C (95°F)
- Storage Range: -40°C (-40°F) to 60°C (140°F)
- Humidity: 10-70% RH non-condensing
- Vibration: < 0.3G over a 10 min period (Holter Recorder and NIBP modules)
- Altitude: -500 to +5000 feet w.r.t. sea level
Note: the instrument's specifications may degrade if operated outside of these environmental specifications
CASSETTE PLAYBACK SPECIFICATIONS:
- Signal Channels: 2 or 3, software selection
- Clock Track: Recorded on channel 3
- Clock Track Frequency: 32Hz
- Scanning Speed: 480 mm/sec
- Rewind Speed: 2000 mm/sec
- Sample Rate: 128 samples/sec nom. (referred to 1mm/sec record speed)
- Sample Signal: Clock track or CD350 tachometer
- Signal Range: ±2.5 mVPP
- Analog Digital Conversion: 8 bits per channel
- Signal Resolution: 19.6 uV per bit
- PC Board Pre-amp Gain: X.5, X1, and X2
- PC Board Amplifier Gain: 256 levels from X0 to X1.99
- PC Board Address Range: 340H - 35FH
- PC Board Data Memory Size: 1MByte
- Data Collection Method: Software - No DMA or interrupt levels needed
- Bezel:
3 ½” molded plastic-Apple platinum color
+12V current:3 amps
-12V current:200ma
+5V current:1 amp - Operating Temperature: +5°C to +45°C
- Trans/Storage Temperature: -40°C to +70°C
SPO2 SPECIFICATIONS:
- Oxygen Saturation Range: 0-100%
- Accuracy: +/- 2% for 70-100%. +/- 3% for 50-69%
- Pulse Rate: 30-250
- Probes: Finger, ear, non-disposable
TEMPERATURE SPECIFICATIONS:
- Range: 80 F to 115 F (26 C to 45 C)
- Accuracy: +/- 0.2 C
- Self-test Between Each Monitor Reading: About 2 seconds
- Probes: Oral and rectal
WEIGHT SPECIFICATIONS:
- Range: 400 lbs (180kg)
- Accuracy: +/- 0.5 lb (0.2kg)
- Platform Size: 12” x 12.25” x 2.25” (30.5x31.1x5.7cm)
HOLTER FLASH RECORDER SPECIFICATIONS:
- Functional Channels: 3
- Resolution: 8-bit (13-bit in the CardioCard™ system)
- Recording: Full disclosure; Minimum Recording Time: 24 hours / 48 hours
- Storage Medium: Removable FLASH memory (Standard PCMCIA-ATA Card)
- Download Interface: PCMCIA card interface
- Pace Detection
- Physical Dimensions: 4.4" x 2.6" x .95"; Weight: 5 ounces with battery
- Enclosure: Molded plastic
- Operating Position: Any orientation
- Operating Temperature: 0 C to 45 C; Non-operating Temperature: -40 C to 60 C;
- Operating Humidity: 10% to 90% (non-condensing)
ELECTRICAL:
- Battery: One AA Alkaline (disposable)
- Connector: 5 or 7 Lead Wires
- Test Jacks: 3 channels
To Contact Support:
Please call 1-866-627-4332 or (315)-676-2346 between 9:00 am - 4:00 pm EST.
Our meeting hours are 9:00 am - 4:00 pm EST.
If you are contacting us after these hours, please
Please call 1-866-627-4332 or (315)-676-2346 between 9:00 am - 4:00 pm EST.
Our meeting hours are 9:00 am - 4:00 pm EST.
If you are contacting us after these hours, please
and we will schedule a time to take care of your needs as quickly as we can.
Thank you for your interest in Nasiff CardioCard® products.
Thank you for your interest in Nasiff CardioCard® products.